Detection of glaucomatous visual field defect using a screening program of Humphrey Field Analyzer.
Chul Hong, Ki Yung Song, Dong Ho Youn, Woo Hyung Park
1Glaucoma Service, Dr. Hong's Eye Clinic, Seoul, Korea.
2Department of Ophthalmology, College of Medicine, Seoul National University, Seoul, Korea.
3Dr. Park's Eye Clinic, Seoul, Korea.
Abstract
To evaluate its clinical efficacy, we report the results of Armaly central field screening test in comparison with those of central 30-2 threshold test of Humphrey Field Analyzer. In 83 cases of normotensives, ocular hypertensives, and glaucoma patients with open-angle or narrow-angle enrolled in this study, a total of 143 eyes were examined. Fifty-four out of 61 patients (88.5%) and 73 out of 81 eyes (90.1%) with visual field defects were detected by the Humphrey Field Analyzer screening test. Eight eyes out of 81(9.9%) with visual field defects detected by the screening test were confirmed as false positive. All points of false positive were solitary, which tended to occur more frequently on superior visual field. A false negative of eight eyes out of 62 (12.9%) detected by the screening test was confirmed by the threshold test, which was more frequent on superior field and shown to be located more peripherally from the fixation point. There was no significant difference in either central sensitivity or age of the false positive and negative.
Key Words: false negative;false postive;glaucomatous visual field;Humphrey Field Analyzer;screening test
TOOLS
METRICS Graph View
  • 2 Crossref
  • 6 Scopus
  • 1,458 View
  • 11 Download
Related articles


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
#1001, Jeokseon Hyundai BD
130, Sajik-ro, Jongno-gu, Seoul 03170, Korea
Tel: +82-2-2271-6788    Fax: +82-2-2277-5194    E-mail: kos@ijpnc.com                

Copyright © 2025 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next