Accurate anterior segment measurements are essential for cataract and refractive surgery planning, as corneal curvature, central corneal thickness (CCT), pupil diameter (PD), and white-to-white (WTW) measurements directly influence intraocular lens (IOL) power calculation and postoperative visual outcomes [
1-
5]. Even small deviations in keratometry can result in clinically meaningful postoperative refractive errors. In addition, parameters such as CCT, PD, and WTW contribute to estimating the effective lens position and determining patient suitability for phakic or premium IOLs, underscoring the need for precise and reproducible biometric measurement.
Recent advancements in optical biometric devices have enabled the measurement of both anterior and posterior corneal curvatures, allowing the acquisition of total keratometry (TK) data. Posterior corneal curvature also contributes meaningfully to overall corneal power, and many studies have shown that disregarding it can result in systematic under- or overestimation, especially in eyes with high astigmatism and prior ophthalmic surgeries. TK is increasingly used in modern IOL formulas to improve accuracy [
4-
6]. The devices evaluated in this study, including Pentacam HR (Oculus), Anterion (Heidelberg Engineering), and IOLMaster 700 (Carl Zeiss Meditec), employ distinct technologies to acquire these important biometric measurements including TK.
Various anterior segment devices are based on different imaging principles and segmentation algorithms. The Pentacam uses a rotating Scheimpflug camera to generate three-dimensional reconstructions of the anterior chamber structure, while the Anterion and IOLMaster 700 are based on swept-source optical coherence tomography (SS-OCT), which provides high-resolution visualization of structures from the anterior corneal surface to the posterior lens [
7,
8]. The IOLMaster 700 additionally combines SS-OCT biometry with a telecentric keratometry algorithm which may induce measurement discrepancies compared with similar OCT-based devices [
9].
Given these inherent differences in optical principles and algorithms, systematic discrepancies in measurements may occur, potentially affecting surgical planning and IOL power calculation [
2,
10]. However, comparative studies among these biometers remain limited to a narrow set of parameters. Some previous studies comparing these devices have reported inconsistent results, and few have simultaneously compared the three biometers. Some found strong correlations among keratometry, CCT, or axial length measurements [
11,
12], but others noted significant differences in TK, PD, or WTW that could influence IOL power estimation [
10,
13-
16]. Therefore, the interchangeability of measurements across devices remains controversial.
As these devices are widely used not only for cataract surgery, but also for refractive surgery planning, and assessment of corneal disease, establishing their reliability and interchangeability in different clinical settings is necessary. Therefore, this study evaluated the inter-device agreement and correlation of anterior keratometry and TK, CCT, PD, and WTW among Anterion, Pentacam, and IOLMaster 700 using both conventional intraclass correlation coefficient (ICC) and Bland-Altman analysis and also applied Deming regression to account for measurement error in both devices.
MATERIALS AND METHODS
Ethics statement
This study was approved by the Institutional Review Board of Yonsei University College of Medicine (No. 4-2025-1106) and was conducted in accordance with the tenets of the Declaration of Helsinki. The requirement for informed consent was waived due to the use of deidentified data and the retrospective nature of the study.
Participants
This retrospective cross-sectional observational study reviewed the medical records of patients who underwent routine preoperative cataract evaluation at Severance Hospital, Yonsei University College of Medicine, between November 2023 and January 2024. Patients who underwent anterior segment measurements using the Anterion, Pentacam, and IOLMaster 700 on the same day were included. A total of 200 eyes from 100 patients were analyzed in this study. After exclusion, 165 eyes from 88 patients were included in the final analysis.
Eyes with a history of ocular trauma or surgery and contact lens wear within 1 week before examination were excluded. Patients with ocular diseases that could affect the measurement quality, including keratoconus, corneal opacity, glaucoma, advanced cataracts impairing fixation, pseudoexfoliation syndrome, uveitis, amblyopia, and retinal disease, were also excluded. Patients with poor image quality or incomplete measurements were excluded from the analysis. All examinations were performed by an experienced examiner on the same day, under standardized mesopic room illumination to minimize pupillary fluctuation, with an interval of 5 minutes between devices. Patients were instructed to blink immediately before image acquisition to ensure a uniform tear film and minimize blinking or eye movement during the measurement.
Parameters
The following parameters were collected for each device. Anterior keratometry values were obtained from each device as flat, steep, and mean anterior keratometry. Total keratometric parameters were obtained as flat, steep, and mean TK (TKm) from the IOLMaster 700, total corneal power (TCP) from the Anterion, and total corneal refractive power (TCRP) from the Pentacam. For simplicity, these are collectively referred to as TK in this manuscript (Anterion TCP, Pentacam TCRP, and IOLMaster TK). The CCT, PD, and WTW were obtained from each device. However, WTW values were unavailable in many cases with Pentacam. Therefore, WTW was analyzed using only the Anterion and IOLMaster 700 devices.
The Pentacam derives anterior curvature and posterior corneal data from Scheimpflug tomography [
17], and anterior keratometry and TCRP were acquired in cataract mode using the central 3.0-mm analysis zone. The Anterion calculates simulated keratometry as the average anterior corneal power over a ray-traced 3.0-mm ring centered on the corneal apex, and TCP incorporates posterior curvature derived from SS-OCT based corneal tomography [
18]. The IOLMaster 700 is also based on SS-OCT and derives anterior keratometry using telecentric keratometry from three concentric rings within the central cornea (approximately 1.5, 2.5, and 3.3 mm in diameter in an average cornea) and reconstructs posterior curvature and corneal thickness from SS-OCT B-scans to calculate TK using a thick lens formula [
19].
CCT was measured using Scheimpflug tomography in the Pentacam and SS-OCT cross-sectional imaging in the Anterion and IOLMaster 700. PD and WTW measurements were derived from anterior segment image analysis using device-specific imaging systems. In Pentacam both parameters were obtained from Scheimpflug-based imaging with automated edge detection of pupil margin and corneal limbus [
20]. Anterion integrates lateral scanning SS-OCT with a 1,300 nm light source and an infrared enface anterior segment camera from which PD and WTW are calculated. IOLMaster 700 employs SS-OCT and an infrared emitting diode based anterior segment camera for automated iris edge detection and horizontal corneal diameter [
21,
22].
Statistical analysis
All statistical analyses were performed using IBM SPSS ver. 27.0 (IBM Corp.). Continuous variables are expressed as mean ± standard deviation. Pairwise comparisons were performed using paired
t-tests or Wilcoxon signed-rank tests when normality assumptions were not met. Comparisons among the three devices were analyzed using repeated-measures analysis of variance with Bonferroni correction for
post-hoc testing. The correlation between the devices was assessed using Pearson correlation coefficients. Agreement was evaluated using the ICC (two-way random effects, single measure, absolute agreement). ICC values were interpreted as follows: <0.50, poor reliability; 0.50 to <0.75, moderate reliability; 0.75 to 0.90, good reliability; and >0.90, excellent reliability [
23].
Bland-Altman plots were constructed using R ver. 4.4.1 (blandr and ggplot2 packages) to calculate the mean bias and 95% limits of agreement (LoA). In addition, Deming regression was applied to account for the measurement error in both devices and provide a more balanced estimation of inter-device relationships. Proportional bias (slope) and constant bias (intercept) were reported with 95% confidence intervals.
To evaluate the potential clinical impact of inter-device differences in TK, IOL power calculations were simulated using Barrett Universal II formula [
24]. TK values from each device were entered with identical biometric inputs (axial length, anterior chamber depth, lens thickness, WTW from the IOLMaster 700), with a target refraction of 0.0 diopter (D), keratometric index of 1.3375, and an A-constant 119.39 (TECNIS ZCB00, Johnson & Johnson Vision). The recommended IOL powers provided by the Barrett Universal II calculator were reported in 0.5 D increments. Pairwise differences in predicted IOL power were calculated for each eye and expressed as mean ± standard deviation. Mean differences were compared using paired
t-tests and absolute differences were further categorized into <0.5 D (no change in lens power), 0.5 ≤ difference < 1.0 D (one-step difference), and ≥1.0 D (two or more differences). Statistical significance was set at
p-value of <0.05.
Results
A total of 200 eyes from 100 patients were initially enrolled, and after exclusion, 165 eyes from 88 patients were included in the final analysis. The mean patient age was 68.6 ± 19.9 years, with 35.7% male patients and 64.3% female patients.
Anterior keratometry
Flat, steep, and mean anterior keratometry values showed small but statistically significant mean differences among the devices (
p ≤ 0.010) (
Table 1). The IOLMaster 700 consistently reported slightly steeper mean keratometry than the Anterion (mean difference [IOLMaster 700 - Anterion], + 0.14 D; 95% confidence interval, 0.096-0.190;
p < 0.001), whereas Pentacam values were intermediate (
Table 2). The correlations across devices were very strong and excellent (r ≥ 0.96, ICC ≥ 0.96) (
Table 3). Bland-Altman analysis revealed mean biases <0.2 D with 95% LoA within approximately ±0.7 D (
Fig. 1A-1C). Deming regression did not show a significant bias for flat keratometry, while comparisons of mean keratometry between the Anterion and IOLMaster 700 demonstrated both constant and proportional biases (
Fig. 2A-2C,
Supplementary Table 1).
Total keratometry
TK exhibited larger discrepancies than anterior keratometry values. For example, the mean TK measured by the IOLMaster 700 was steeper than Anterion (mean difference [IOLMaster 700 - Anterion], 0.71 D;
p < 0.001) and Pentacam (mean difference [IOLMaster 700 - Pentacam], 0.39 D;
p < 0.001) (
Table 2). Pentacam TCRP values were intermediate. Although the correlations remained high (r ≥ 0.95), the ICC values were slightly lower (0.87-0.93) (
Table 3). Bland-Altman plots showed LoA exceeding ±0.5 D and within ±0.9 D in several comparisons (
Fig. 1D-1F). Deming regression identified proportional bias in multiple TK comparisons, indicating increasing divergence at steeper curvatures and constant bias in Anterion (mean TCP) vs. Pentacam (mean TCRP) (
Fig. 2D-2F,
Supplementary Table 1).
Central corneal thickness
The CCT showed statistically significant mean differences among the devices (
p < 0.001) (
Table 1). Pentacam measured thicker values than both Anterion (mean difference [Pentacam - Anterion], 8.0 μm;
p < 0.001) and IOLMaster 700 (mean difference [Pentacam - IOLMaster 700], 6.9 μm;
p < 0.001). No statistically significant difference was observed between Anterion and IOLMaster 700 (
p = 0.066) (
Table 2). Despite these systematic differences, the LoA remained relatively narrow across all pairwise comparisons, and the correlations were strong (r ≥ 0.98, ICC ≥ 0.96) (
Fig. 3A-3C,
Table 3). Deming regression confirmed a constant bias between Pentacam and Anterion and both constant and proportional biases between Pentacam and IOLMaster 700, consistent with the mean differences (
Table 2,
Fig. 4A-4C,
Supplementary Table 2).
Pupil diameter
PD demonstrated the lowest consistency among the parameters. The Pentacam measured pupils that were nearly 1.0 mm smaller than those measured by the Anterion and IOLMaster 700 (
p < 0.001), whereas the latter two devices produced comparable results (
Table 2). Bland-Altman plots revealed wide LoA and the correlations ranged from poor to moderate (r = 0.59-0.80, ICC = 0.41-0.80) (
Fig. 3D-3F,
Table 3). Deming regression showed a constant bias only in the Pentacam versus IOLMaster 700 comparison, while no significant bias was found between Pentacam and Anterion or between the two OCT-based devices (
Fig. 4D-4F,
Supplementary Table 2).
White-to-white
WTW also differed significantly between the devices. The IOLMaster 700 reported slightly larger values than the Anterion (mean difference [IOLMaster 700 - Anterion], 0.19 mm;
p < 0.001) (
Table 2). The correlation was moderate (r = 0.77, ICC = 0.68) (
Table 3). Bland-Altman analysis showed a mean bias of 0.19 mm with LoA ranging from -0.32 to 0.71 mm (
Fig. 3G). Deming regression did not confirm statistically significant slope or intercept bias (
Fig. 4G,
Supplementary Table 2).
Clinical impact of inter-device TK discrepancies on IOL power calculation
Simulated IOL power calculations using the Barrett Universal II formula revealed significant inter-device differences. Simulated IOL power calculated using the Anterion was significantly higher than that calculated using IOLMaster 700 (mean difference [Anterion - IOLMaster 700], 0.87 D;
p < 0.001), with 63.6% of eyes demonstrating an absolute difference of ≥1.0 D. Simulated IOL power calculated using the Pentacam was significantly higher than that calculated using IOLMaster 700 (mean difference [Pentacam - IOLMaster 700], 0.48 D;
p < 0.001), with 44.1% of eyes demonstrating an absolute difference of 0.5 to <1.0 D. Simulated IOL power calculated using the Pentacam was significantly lower than that calculated using Anterion (mean difference [Pentacam - Anterion], -0.39 D;
p < 0.001), with 44.8% of eyes demonstrating an absolute difference of 0.5 to <1.0 D (
Table 4).
Discussion
This study demonstrated that the degree of inter-device agreement among the Anterion, Pentacam, and IOLMaster 700 differed depending on the parameters being measured. While simulated keratometry (SimK) and CCT showed excellent ICC agreement and small mean differences, clinically relevant biases in some parameters were noted in the Bland-Altman and Deming regression analyses. Overall, these findings highlight the importance of distinguishing statistical reliability from clinical equivalence when assessing anterior segment measurements and considering whether the values can be used interchangeably. Because these three devices are the most commonly used biometric platforms in clinical settings, these findings have important implications for proper device selection and interpretation of biometric data from different platforms.
Anterior keratometry showed very strong reliability, with the highest ICCs and smallest mean differences across all devices. These results align with previous studies reporting strong agreement in anterior corneal curvature between Pentacam and SS-OCT based biometers [
12,
25,
26]. However, the LoA occasionally exceeded the ±0.50 D range, which is commonly regarded as the clinically acceptable range for IOL power calculations [
27,
28]. The keratometric values measured by the IOLMaster 700 tended to be steeper, followed by those of the Pentacam and Anterion. Deming regression identified both constant and proportional bias between the Anterion and IOLMaster 700, suggesting that the difference could increase with steeper corneal curvature. Although SimK showed greater consistency than TK, full interchangeability across devices remains limited.
In contrast, TK showed greater inter-device variability than SimK. Despite the strong correlations, the lower ICC results and wider LoA indicate that the measurements are not numerically interchangeable. Previous studies have similarly concluded that TK values are not interchangeable among devices [
29]. Deming regression revealed a larger constant and proportional bias between several pairs of devices and also suggested that discrepancies increased in steeper corneas and that their magnitude was not constant across the curvature range. Measurement zone differences and different types of measurement algorithms may partly contribute to the disparities among devices.
These differences are clinically important because modern IOL power calculation formulas increasingly incorporate TK [
30]. Our simulation analysis in
Table 4 demonstrated that such differences in TK could translate into clinically meaningful variability in IOL power selection. Because commercially available IOL powers are typically offered in 0.5 D increments, an interdevice difference of ≥1.0 D in predicted IOL power corresponds to at least a two-step difference and may be clinically meaningful. Such discrepancies may contribute to refractive prediction errors, particularly in eyes with high astigmatism or irregular corneal shapes. Therefore, TK values should be interpreted in the context of each device’s algorithm and used consistently with the same device throughout preoperative assessment.
CCT measurements also showed systematic differences between the devices. The Pentacam consistently reported thicker values than both Anterion and IOLMaster 700, whereas no statistically significant difference was observed between the Anterion and IOLMaster 700. Some prior studies attributed such discrepancies to differences in the assumed refractive indices and segmentation algorithms [
31,
32]. Despite these systematic biases, the correlation coefficients in our study were nearly perfect (r = 0.99, ICC > 0.96), indicating that CCT measurements are highly reliable when used longitudinally with the same device but should not be directly compared across different platforms. Clinically, this distinction is also important in refractive surgery planning and corneal disease monitoring, in which absolute corneal thickness plays a critical role.
Among the evaluated parameters, PD showed the lowest inter-device agreement. The Pentacam consistently measured smaller pupils than the two SS-OCT based devices, although Deming regression revealed a constant bias only between the Pentacam and the IOLMaster 700. No significant bias was observed between the Anterion and IOLMaster 700 measurements. These discrepancies between Scheimpflug and SS-OCT based systems could be due to variations in illumination conditions and pupil-edge detection algorithms [
33]. Given that pupil size directly affects the optical performance of multifocal and extended depth of focus IOLs, consistent device use and cross-checking between devices are recommended for PD assessment in preoperative IOL planning [
34].
WTW measurements showed only moderate agreement (r = 0.77, ICC = 0.68). The Anterion consistently reported smaller WTW values than the IOLMaster 700, which is in agreement with the findings of previous comparative studies [
22]. These disparities are probably caused by variations in the imaging light sources and boundary detection algorithms, which influence the recognition of the limbal edge. In addition, peripheral corneal irregularities may contribute to inter-device variability [
35]. Although the regression coefficients were not statistically significant, the mean difference of approximately 0.2 mm is potentially clinically relevant, as WTW is a critical parameter in sizing intraocular implants, such as phakic IOLs or capsular tension rings [
36]. Even small inter-device discrepancies can influence surgical planning and should be interpreted with caution.
In addition to conventional ICC and Bland-Altman analyses, Deming regression was performed to investigate systematic bias across devices. By accounting for the measurement error in both axes, Deming regression provides additional insight into the direction and magnitude of inter- device discrepancies. Both constant and proportional biases were identified across several parameters, most notably in the TK and corneal thickness measurements. These systematic discrepancies were observed across multiple device pair comparisons, not limited to a single device pair comparison, indicating that they are unlikely to be attributable to random variation but likely reflect the fundamental differences in optical principles and segmentation algorithms. Although conversion equations derived from regression may help clarify the numerical and intuitive relationships between devices, they should be interpreted with caution, as they are influenced by population-level trends and may not reflect individual variability.
A major strength of this study is that it represents the first comparative analysis in South Korea to simultaneously evaluate five major anterior segment measurements, including both anterior and TK, as well as CCT, PD, and WTW obtained from three of the most commonly used biometers. In addition, our sample size of 165 eyes exceeds that of most previous domestic reports, which typically included fewer than 100 eyes, thereby enhancing the statistical robustness and generalizability of our findings.
This study has several limitations. This was a retrospective, single-center study, involving only cataract surgery candidates, which may limit generalizability. The comparison of IOL power calculations based on TK from each device was a simulation rather than an analysis of actual postoperative refractive outcomes. Therefore, actual clinical impact should be interpreted cautiously and future prospective studies incorporating postoperative refractive outcomes are warranted to validate the clinical implications observed in this simulation analysis. In addition, all measurements were obtained once by a single experienced examiner, and inter-visit repeatability and inter-operator variability were not assessed. Nevertheless, the relatively large sample size supports the reliability of our findings.
In conclusion, although anterior keratometry and CCT measurements showed strong statistical agreement among the three devices, small but consistent biases indicated that they were not fully interchangeable. Large systematic disparities were observed in TK, PD, and WTW, underscoring the necessity of consistent device use, cross-checking among devices or the application of validated conversion when these parameters are applied clinically. Deming regression provided complementary insights by showing constant and proportional biases, emphasizing that a narrow LoA and strong statistical correlation do not guarantee clinical equivalence.