Jung, Lee, and Kim: Comparative Analysis of Anterior Segment Biometry: Agreement and Correlation between Swept-Source Optical Coherence Tomography and Scheimpflug-Based Platforms

Abstract

Purpose

To evaluate the inter-device agreement of keratometry, central corneal thickness (CCT), pupil diameter (PD), and white-to-white (WTW) measurements among three anterior segment biometers: Anterion (Heidelberg Engineering), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec).

Methods

This retrospective study analyzed 165 eyes of 88 patients who underwent measurements using all three devices on the same day under standardized mesopic illumination. Parameters included anterior and total keratometry, CCT, PD, and WTW. Agreement was assessed using repeated-measures ANOVA, intraclass correlation coefficient (ICC), Bland-Altman analysis, and Deming regression.

Results

Anterior keratometry (simulated keratometry [SimK]) and CCT showed excellent agreement (ICC ≥ 0.96) with minimal bias (<0.2 diopters and approximately 8 μm, respectively). The IOLMaster 700 measured slightly steeper keratometry than the Anterion, whereas the Pentacam yielded thicker CCT values than both IOLMaster 700 and Anterion. Total keratometry exhibited larger discrepancies than anterior keratometry. The Anterion reported smaller WTW measurements than the IOLMaster 700. Deming regression revealed proportional bias in total keratometry and constant bias in CCT and PD across the three devices.

Conclusions

Although SimK and CCT showed a strong statistical agreement, they were not found to be fully interchangeable. The need for consistent use of devices or validated conversion methods in clinical settings is emphasized by the significant differences in total keratometry, PD, and WTW. These findings are clinically relevant for proper cataract surgery, refractive surgery planning, and diagnosis of corneal diseases.

Accurate anterior segment measurements are essential for cataract and refractive surgery planning, as corneal curvature, central corneal thickness (CCT), pupil diameter (PD), and white-to-white (WTW) measurements directly influence intraocular lens (IOL) power calculation and postoperative visual outcomes [1-5]. Even small deviations in keratometry can result in clinically meaningful postoperative refractive errors. In addition, parameters such as CCT, PD, and WTW contribute to estimating the effective lens position and determining patient suitability for phakic or premium IOLs, underscoring the need for precise and reproducible biometric measurement.
Recent advancements in optical biometric devices have enabled the measurement of both anterior and posterior corneal curvatures, allowing the acquisition of total keratometry (TK) data. Posterior corneal curvature also contributes meaningfully to overall corneal power, and many studies have shown that disregarding it can result in systematic under- or overestimation, especially in eyes with high astigmatism and prior ophthalmic surgeries. TK is increasingly used in modern IOL formulas to improve accuracy [4-6]. The devices evaluated in this study, including Pentacam HR (Oculus), Anterion (Heidelberg Engineering), and IOLMaster 700 (Carl Zeiss Meditec), employ distinct technologies to acquire these important biometric measurements including TK.
Various anterior segment devices are based on different imaging principles and segmentation algorithms. The Pentacam uses a rotating Scheimpflug camera to generate three-dimensional reconstructions of the anterior chamber structure, while the Anterion and IOLMaster 700 are based on swept-source optical coherence tomography (SS-OCT), which provides high-resolution visualization of structures from the anterior corneal surface to the posterior lens [7,8]. The IOLMaster 700 additionally combines SS-OCT biometry with a telecentric keratometry algorithm which may induce measurement discrepancies compared with similar OCT-based devices [9].
Given these inherent differences in optical principles and algorithms, systematic discrepancies in measurements may occur, potentially affecting surgical planning and IOL power calculation [2,10]. However, comparative studies among these biometers remain limited to a narrow set of parameters. Some previous studies comparing these devices have reported inconsistent results, and few have simultaneously compared the three biometers. Some found strong correlations among keratometry, CCT, or axial length measurements [11,12], but others noted significant differences in TK, PD, or WTW that could influence IOL power estimation [10,13-16]. Therefore, the interchangeability of measurements across devices remains controversial.
As these devices are widely used not only for cataract surgery, but also for refractive surgery planning, and assessment of corneal disease, establishing their reliability and interchangeability in different clinical settings is necessary. Therefore, this study evaluated the inter-device agreement and correlation of anterior keratometry and TK, CCT, PD, and WTW among Anterion, Pentacam, and IOLMaster 700 using both conventional intraclass correlation coefficient (ICC) and Bland-Altman analysis and also applied Deming regression to account for measurement error in both devices.

MATERIALS AND METHODS

Ethics statement

This study was approved by the Institutional Review Board of Yonsei University College of Medicine (No. 4-2025-1106) and was conducted in accordance with the tenets of the Declaration of Helsinki. The requirement for informed consent was waived due to the use of deidentified data and the retrospective nature of the study.

Participants

This retrospective cross-sectional observational study reviewed the medical records of patients who underwent routine preoperative cataract evaluation at Severance Hospital, Yonsei University College of Medicine, between November 2023 and January 2024. Patients who underwent anterior segment measurements using the Anterion, Pentacam, and IOLMaster 700 on the same day were included. A total of 200 eyes from 100 patients were analyzed in this study. After exclusion, 165 eyes from 88 patients were included in the final analysis.
Eyes with a history of ocular trauma or surgery and contact lens wear within 1 week before examination were excluded. Patients with ocular diseases that could affect the measurement quality, including keratoconus, corneal opacity, glaucoma, advanced cataracts impairing fixation, pseudoexfoliation syndrome, uveitis, amblyopia, and retinal disease, were also excluded. Patients with poor image quality or incomplete measurements were excluded from the analysis. All examinations were performed by an experienced examiner on the same day, under standardized mesopic room illumination to minimize pupillary fluctuation, with an interval of 5 minutes between devices. Patients were instructed to blink immediately before image acquisition to ensure a uniform tear film and minimize blinking or eye movement during the measurement.

Parameters

The following parameters were collected for each device. Anterior keratometry values were obtained from each device as flat, steep, and mean anterior keratometry. Total keratometric parameters were obtained as flat, steep, and mean TK (TKm) from the IOLMaster 700, total corneal power (TCP) from the Anterion, and total corneal refractive power (TCRP) from the Pentacam. For simplicity, these are collectively referred to as TK in this manuscript (Anterion TCP, Pentacam TCRP, and IOLMaster TK). The CCT, PD, and WTW were obtained from each device. However, WTW values were unavailable in many cases with Pentacam. Therefore, WTW was analyzed using only the Anterion and IOLMaster 700 devices.
The Pentacam derives anterior curvature and posterior corneal data from Scheimpflug tomography [17], and anterior keratometry and TCRP were acquired in cataract mode using the central 3.0-mm analysis zone. The Anterion calculates simulated keratometry as the average anterior corneal power over a ray-traced 3.0-mm ring centered on the corneal apex, and TCP incorporates posterior curvature derived from SS-OCT based corneal tomography [18]. The IOLMaster 700 is also based on SS-OCT and derives anterior keratometry using telecentric keratometry from three concentric rings within the central cornea (approximately 1.5, 2.5, and 3.3 mm in diameter in an average cornea) and reconstructs posterior curvature and corneal thickness from SS-OCT B-scans to calculate TK using a thick lens formula [19].
CCT was measured using Scheimpflug tomography in the Pentacam and SS-OCT cross-sectional imaging in the Anterion and IOLMaster 700. PD and WTW measurements were derived from anterior segment image analysis using device-specific imaging systems. In Pentacam both parameters were obtained from Scheimpflug-based imaging with automated edge detection of pupil margin and corneal limbus [20]. Anterion integrates lateral scanning SS-OCT with a 1,300 nm light source and an infrared enface anterior segment camera from which PD and WTW are calculated. IOLMaster 700 employs SS-OCT and an infrared emitting diode based anterior segment camera for automated iris edge detection and horizontal corneal diameter [21,22].

Statistical analysis

All statistical analyses were performed using IBM SPSS ver. 27.0 (IBM Corp.). Continuous variables are expressed as mean ± standard deviation. Pairwise comparisons were performed using paired t-tests or Wilcoxon signed-rank tests when normality assumptions were not met. Comparisons among the three devices were analyzed using repeated-measures analysis of variance with Bonferroni correction for post-hoc testing. The correlation between the devices was assessed using Pearson correlation coefficients. Agreement was evaluated using the ICC (two-way random effects, single measure, absolute agreement). ICC values were interpreted as follows: <0.50, poor reliability; 0.50 to <0.75, moderate reliability; 0.75 to 0.90, good reliability; and >0.90, excellent reliability [23].
Bland-Altman plots were constructed using R ver. 4.4.1 (blandr and ggplot2 packages) to calculate the mean bias and 95% limits of agreement (LoA). In addition, Deming regression was applied to account for the measurement error in both devices and provide a more balanced estimation of inter-device relationships. Proportional bias (slope) and constant bias (intercept) were reported with 95% confidence intervals.
To evaluate the potential clinical impact of inter-device differences in TK, IOL power calculations were simulated using Barrett Universal II formula [24]. TK values from each device were entered with identical biometric inputs (axial length, anterior chamber depth, lens thickness, WTW from the IOLMaster 700), with a target refraction of 0.0 diopter (D), keratometric index of 1.3375, and an A-constant 119.39 (TECNIS ZCB00, Johnson & Johnson Vision). The recommended IOL powers provided by the Barrett Universal II calculator were reported in 0.5 D increments. Pairwise differences in predicted IOL power were calculated for each eye and expressed as mean ± standard deviation. Mean differences were compared using paired t-tests and absolute differences were further categorized into <0.5 D (no change in lens power), 0.5 ≤ difference < 1.0 D (one-step difference), and ≥1.0 D (two or more differences). Statistical significance was set at p-value of <0.05.

Results

A total of 200 eyes from 100 patients were initially enrolled, and after exclusion, 165 eyes from 88 patients were included in the final analysis. The mean patient age was 68.6 ± 19.9 years, with 35.7% male patients and 64.3% female patients.

Anterior keratometry

Flat, steep, and mean anterior keratometry values showed small but statistically significant mean differences among the devices (p ≤ 0.010) (Table 1). The IOLMaster 700 consistently reported slightly steeper mean keratometry than the Anterion (mean difference [IOLMaster 700 - Anterion], + 0.14 D; 95% confidence interval, 0.096-0.190; p < 0.001), whereas Pentacam values were intermediate (Table 2). The correlations across devices were very strong and excellent (r ≥ 0.96, ICC ≥ 0.96) (Table 3). Bland-Altman analysis revealed mean biases <0.2 D with 95% LoA within approximately ±0.7 D (Fig. 1A-1C). Deming regression did not show a significant bias for flat keratometry, while comparisons of mean keratometry between the Anterion and IOLMaster 700 demonstrated both constant and proportional biases (Fig. 2A-2C, Supplementary Table 1).

Total keratometry

TK exhibited larger discrepancies than anterior keratometry values. For example, the mean TK measured by the IOLMaster 700 was steeper than Anterion (mean difference [IOLMaster 700 - Anterion], 0.71 D; p < 0.001) and Pentacam (mean difference [IOLMaster 700 - Pentacam], 0.39 D; p < 0.001) (Table 2). Pentacam TCRP values were intermediate. Although the correlations remained high (r ≥ 0.95), the ICC values were slightly lower (0.87-0.93) (Table 3). Bland-Altman plots showed LoA exceeding ±0.5 D and within ±0.9 D in several comparisons (Fig. 1D-1F). Deming regression identified proportional bias in multiple TK comparisons, indicating increasing divergence at steeper curvatures and constant bias in Anterion (mean TCP) vs. Pentacam (mean TCRP) (Fig. 2D-2F, Supplementary Table 1).

Central corneal thickness

The CCT showed statistically significant mean differences among the devices (p < 0.001) (Table 1). Pentacam measured thicker values than both Anterion (mean difference [Pentacam - Anterion], 8.0 μm; p < 0.001) and IOLMaster 700 (mean difference [Pentacam - IOLMaster 700], 6.9 μm; p < 0.001). No statistically significant difference was observed between Anterion and IOLMaster 700 (p = 0.066) (Table 2). Despite these systematic differences, the LoA remained relatively narrow across all pairwise comparisons, and the correlations were strong (r ≥ 0.98, ICC ≥ 0.96) (Fig. 3A-3C, Table 3). Deming regression confirmed a constant bias between Pentacam and Anterion and both constant and proportional biases between Pentacam and IOLMaster 700, consistent with the mean differences (Table 2, Fig. 4A-4C, Supplementary Table 2).

Pupil diameter

PD demonstrated the lowest consistency among the parameters. The Pentacam measured pupils that were nearly 1.0 mm smaller than those measured by the Anterion and IOLMaster 700 (p < 0.001), whereas the latter two devices produced comparable results (Table 2). Bland-Altman plots revealed wide LoA and the correlations ranged from poor to moderate (r = 0.59-0.80, ICC = 0.41-0.80) (Fig. 3D-3F, Table 3). Deming regression showed a constant bias only in the Pentacam versus IOLMaster 700 comparison, while no significant bias was found between Pentacam and Anterion or between the two OCT-based devices (Fig. 4D-4F, Supplementary Table 2).

White-to-white

WTW also differed significantly between the devices. The IOLMaster 700 reported slightly larger values than the Anterion (mean difference [IOLMaster 700 - Anterion], 0.19 mm; p < 0.001) (Table 2). The correlation was moderate (r = 0.77, ICC = 0.68) (Table 3). Bland-Altman analysis showed a mean bias of 0.19 mm with LoA ranging from -0.32 to 0.71 mm (Fig. 3G). Deming regression did not confirm statistically significant slope or intercept bias (Fig. 4G, Supplementary Table 2).

Clinical impact of inter-device TK discrepancies on IOL power calculation

Simulated IOL power calculations using the Barrett Universal II formula revealed significant inter-device differences. Simulated IOL power calculated using the Anterion was significantly higher than that calculated using IOLMaster 700 (mean difference [Anterion - IOLMaster 700], 0.87 D; p < 0.001), with 63.6% of eyes demonstrating an absolute difference of ≥1.0 D. Simulated IOL power calculated using the Pentacam was significantly higher than that calculated using IOLMaster 700 (mean difference [Pentacam - IOLMaster 700], 0.48 D; p < 0.001), with 44.1% of eyes demonstrating an absolute difference of 0.5 to <1.0 D. Simulated IOL power calculated using the Pentacam was significantly lower than that calculated using Anterion (mean difference [Pentacam - Anterion], -0.39 D; p < 0.001), with 44.8% of eyes demonstrating an absolute difference of 0.5 to <1.0 D (Table 4).

Discussion

This study demonstrated that the degree of inter-device agreement among the Anterion, Pentacam, and IOLMaster 700 differed depending on the parameters being measured. While simulated keratometry (SimK) and CCT showed excellent ICC agreement and small mean differences, clinically relevant biases in some parameters were noted in the Bland-Altman and Deming regression analyses. Overall, these findings highlight the importance of distinguishing statistical reliability from clinical equivalence when assessing anterior segment measurements and considering whether the values can be used interchangeably. Because these three devices are the most commonly used biometric platforms in clinical settings, these findings have important implications for proper device selection and interpretation of biometric data from different platforms.
Anterior keratometry showed very strong reliability, with the highest ICCs and smallest mean differences across all devices. These results align with previous studies reporting strong agreement in anterior corneal curvature between Pentacam and SS-OCT based biometers [12,25,26]. However, the LoA occasionally exceeded the ±0.50 D range, which is commonly regarded as the clinically acceptable range for IOL power calculations [27,28]. The keratometric values measured by the IOLMaster 700 tended to be steeper, followed by those of the Pentacam and Anterion. Deming regression identified both constant and proportional bias between the Anterion and IOLMaster 700, suggesting that the difference could increase with steeper corneal curvature. Although SimK showed greater consistency than TK, full interchangeability across devices remains limited.
In contrast, TK showed greater inter-device variability than SimK. Despite the strong correlations, the lower ICC results and wider LoA indicate that the measurements are not numerically interchangeable. Previous studies have similarly concluded that TK values are not interchangeable among devices [29]. Deming regression revealed a larger constant and proportional bias between several pairs of devices and also suggested that discrepancies increased in steeper corneas and that their magnitude was not constant across the curvature range. Measurement zone differences and different types of measurement algorithms may partly contribute to the disparities among devices.
These differences are clinically important because modern IOL power calculation formulas increasingly incorporate TK [30]. Our simulation analysis in Table 4 demonstrated that such differences in TK could translate into clinically meaningful variability in IOL power selection. Because commercially available IOL powers are typically offered in 0.5 D increments, an interdevice difference of ≥1.0 D in predicted IOL power corresponds to at least a two-step difference and may be clinically meaningful. Such discrepancies may contribute to refractive prediction errors, particularly in eyes with high astigmatism or irregular corneal shapes. Therefore, TK values should be interpreted in the context of each device’s algorithm and used consistently with the same device throughout preoperative assessment.
CCT measurements also showed systematic differences between the devices. The Pentacam consistently reported thicker values than both Anterion and IOLMaster 700, whereas no statistically significant difference was observed between the Anterion and IOLMaster 700. Some prior studies attributed such discrepancies to differences in the assumed refractive indices and segmentation algorithms [31,32]. Despite these systematic biases, the correlation coefficients in our study were nearly perfect (r = 0.99, ICC > 0.96), indicating that CCT measurements are highly reliable when used longitudinally with the same device but should not be directly compared across different platforms. Clinically, this distinction is also important in refractive surgery planning and corneal disease monitoring, in which absolute corneal thickness plays a critical role.
Among the evaluated parameters, PD showed the lowest inter-device agreement. The Pentacam consistently measured smaller pupils than the two SS-OCT based devices, although Deming regression revealed a constant bias only between the Pentacam and the IOLMaster 700. No significant bias was observed between the Anterion and IOLMaster 700 measurements. These discrepancies between Scheimpflug and SS-OCT based systems could be due to variations in illumination conditions and pupil-edge detection algorithms [33]. Given that pupil size directly affects the optical performance of multifocal and extended depth of focus IOLs, consistent device use and cross-checking between devices are recommended for PD assessment in preoperative IOL planning [34].
WTW measurements showed only moderate agreement (r = 0.77, ICC = 0.68). The Anterion consistently reported smaller WTW values than the IOLMaster 700, which is in agreement with the findings of previous comparative studies [22]. These disparities are probably caused by variations in the imaging light sources and boundary detection algorithms, which influence the recognition of the limbal edge. In addition, peripheral corneal irregularities may contribute to inter-device variability [35]. Although the regression coefficients were not statistically significant, the mean difference of approximately 0.2 mm is potentially clinically relevant, as WTW is a critical parameter in sizing intraocular implants, such as phakic IOLs or capsular tension rings [36]. Even small inter-device discrepancies can influence surgical planning and should be interpreted with caution.
In addition to conventional ICC and Bland-Altman analyses, Deming regression was performed to investigate systematic bias across devices. By accounting for the measurement error in both axes, Deming regression provides additional insight into the direction and magnitude of inter- device discrepancies. Both constant and proportional biases were identified across several parameters, most notably in the TK and corneal thickness measurements. These systematic discrepancies were observed across multiple device pair comparisons, not limited to a single device pair comparison, indicating that they are unlikely to be attributable to random variation but likely reflect the fundamental differences in optical principles and segmentation algorithms. Although conversion equations derived from regression may help clarify the numerical and intuitive relationships between devices, they should be interpreted with caution, as they are influenced by population-level trends and may not reflect individual variability.
A major strength of this study is that it represents the first comparative analysis in South Korea to simultaneously evaluate five major anterior segment measurements, including both anterior and TK, as well as CCT, PD, and WTW obtained from three of the most commonly used biometers. In addition, our sample size of 165 eyes exceeds that of most previous domestic reports, which typically included fewer than 100 eyes, thereby enhancing the statistical robustness and generalizability of our findings.
This study has several limitations. This was a retrospective, single-center study, involving only cataract surgery candidates, which may limit generalizability. The comparison of IOL power calculations based on TK from each device was a simulation rather than an analysis of actual postoperative refractive outcomes. Therefore, actual clinical impact should be interpreted cautiously and future prospective studies incorporating postoperative refractive outcomes are warranted to validate the clinical implications observed in this simulation analysis. In addition, all measurements were obtained once by a single experienced examiner, and inter-visit repeatability and inter-operator variability were not assessed. Nevertheless, the relatively large sample size supports the reliability of our findings.
In conclusion, although anterior keratometry and CCT measurements showed strong statistical agreement among the three devices, small but consistent biases indicated that they were not fully interchangeable. Large systematic disparities were observed in TK, PD, and WTW, underscoring the necessity of consistent device use, cross-checking among devices or the application of validated conversion when these parameters are applied clinically. Deming regression provided complementary insights by showing constant and proportional biases, emphasizing that a narrow LoA and strong statistical correlation do not guarantee clinical equivalence.

Notes

Conflicts of Interest

Tae-im Kim is a section editor of this journal, but was not involved in the peer reviewer selection, evaluation, or decision process of this article. No other potential conflict of interest relevant to this article was reported.

Acknowledgements

None.

Funding

This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean Ministry of Science and ICT (No. RS-2024-00352929).

Supplementary Materials

Supplementary Table 1. Deming regression conversion formulas for keratometric parameters
Supplementary Table 2. Deming regression conversion formulas for CCT, PD, and WTW parameters
kjo-2025-0202-Supplementary-Table.pdf

References

1. Gim Y, Jun RM, Han KE. Agreement between Scheimpflug camera and the swept-source optical coherence tomography measurements in keratometry and higher-order aberrations. Korean J Ophthalmol 2021;35:337-48.
crossref pmid pmc pdf
2. Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols: the better way to design and report clinical trials. Ophthalmology 2021;128:e115-20.
crossref pmid
3. Kim KY, Choi GS, Kang MS, Kim US. Comparison study of the axial length measured using the new swept-source optical coherence tomography ANTERION and the partial coherence interferometry IOL Master. PLoS One 2020;15:e0244590..
crossref pmid pmc
4. Koch DD, Hill W, Abulafia A, Wang L. Pursuing perfection in intraocular lens calculations: I. logical approach for classifying IOL calculation formulas. J Cataract Refract Surg 2017;43:717-8.
crossref pmid
5. LaHood BR, Goggin M. Measurement of posterior corneal astigmatism by the IOLMaster 700. J Refract Surg 2018;34:331-6.
crossref pmid
6. Fabian E, Wehner W. Prediction accuracy of total keratometry compared to standard keratometry using different intraocular lens power formulas. J Refract Surg 2019;35:362-8.
crossref pmid
7. Motlagh MN, Moshirfar M, Murri MS, et al. Pentacam® corneal tomography for screening of refractive surgery candidates: a review of the literature, part I. Med Hypothesis Discov Innov Ophthalmol 2019;8:177-203.
pmid pmc
8. Park HJ, Lee H, Woo YJ, et al. Comparison of the astigmatic power of toric intraocular lenses using three toric calculators. Yonsei Med J 2015;56:1097-105.
crossref pmid pmc
9. Pathak M, Sahu V, Kumar A, et al. Current concepts and recent updates of optical biometry: a comprehensive review. Clin Ophthalmol 2024;18:1191-206.
crossref pmid pmc pdf
10. Ruiz-Mesa R, Aguilar-Corcoles S, Montes-Mico R, Tana- Rivero P. Ocular biometric repeatability using a new high-resolution swept-source optical coherence tomographer. Expert Rev Med Devices 2020;17:591-7.
crossref pmid
11. Ostadi-Moghaddam H, Shandiz JH, Validad MH, et al. Repeatability and agreement of Anterion with Pentacam HR and Orbscan II in corneal parameters after photorefractive keratectomy. Dubai Med J 2024;7:177-92.
crossref pdf
12. Anwar M, Buhren J, Kook D, et al. Comparison of Anterion, MS-39, Pentacam, and Sirius in corneal assessment prior to refractive surgery. Clin Ophthalmol 2025;19:3487-95.
crossref pmid pdf
13. Xu J, Zhang L, Mo E, et al. The effect of corneal power on the accuracy of 14 IOL power formulas. BMC Ophthalmol 2024;24:126.
crossref pmid pmc pdf
14. Liu Q, Yang W, Li D, et al. Total and simulated keratometry measurements using IOLMaster 700 and Pentacam AXL after small incision lenticule extraction. Int Ophthalmol 2024;44:237.
crossref pmid pdf
15. Budiman B, Knoch AM, Boesoirie SF, et al. Agreement between IOLMaster 700 and Pentacam AXL for IOL power measurement in patients with high myopia. Indian J Ophthalmol 2024;72:1021-5.
crossref pmid pmc
16. Jin A, Han X, Zhang J, et al. Agreement of total keratometry and posterior keratometry among IOLMaster 700, CASIA2, and Pentacam. Transl Vis Sci Technol 2023;12:13.
crossref
17. Sinjab MM. Step by step: reading Pentacam topography: basics and clinical applications. 3rd ed. JP Medical Ltd; 2020.
18. McLintock C, Niyazmand H, Seo S, et al. Agreement between 2 SS-OCT biometry devices. J Cataract Refract Surg 2022;48:1107-12.
crossref pmid
19. Savini G, Taroni L, Schiano-Lomoriello D, Hoffer KJ. Repeatability of total keratometry and standard keratometry by the IOLMaster 700 and comparison to total corneal astigmatism by Scheimpf lug imaging. Eye (Lond) 2021;35:307-15.
crossref pmid pdf
20. Dominguez-Vicent A, Monsalvez-Romin D, Aguila-Carrasco AJ, et al. Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpf lug imaging devices. Arq Bras Oftalmol 2014;77:233-7.
pmid
21. Romero D, Toledano R, Carceles A, et al. Comparison meta- analysis of two swept-source optical coherence biometers. Indian J Ophthalmol 2026;74:23-8.
crossref pmid
22. Salouti R, Azizi A, Hashemian H, et al. A comparison of white-to-white measurements obtained by Anterion ASOCT versus three optical devices in refractive surgery candidates. Eur J Ophthalmol 2025;35:902-9.
crossref pmid pdf
23. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med 2016;15:155-63.
crossref pmid pmc
24. Barrett GD. An improved universal theoretical formula for intraocular lens power prediction. J Cataract Refract Surg 1993;19:713-20.
crossref pmid
25. Wang ZY, Yang WL, Li DJ, et al. Comparison of biometry with the Pentacam AXL, IOLMaster 700 and IOLMaster 500 in cataract patients. Zhonghua Yan Ke Za Zhi 2019;55:515-21.
pmid
26. Kim BY, Jun I. Comparison of anterior segment measurements with a new multifunctional unit and five other devices. Korean J Ophthalmol 2022;36:338-49.
crossref pmid pmc pdf
27. Olsen T. Calculation of intraocular lens power: a review. Acta Ophthalmol Scand 2007;85:472-85.
crossref pmid
28. Shirayama M, Wang L, Weikert MP, Koch DD. Comparison of corneal powers obtained from 4 different devices. Am J Ophthalmol 2009;148:528-35. (e1.
crossref pmid
29. Muzyka-Wozniak M, Oleszko A, Grzybowski A. Measurements of anterior and posterior corneal curvatures with OCT and Scheimpflug biometers in patients with low total corneal astigmatism. J Clin Med 2022;11:6921.
crossref pmid pmc
30. Wei L, Cheng K, He W, et al. Application of total keratometry in ten intraocular lens power calculation formulas in highly myopic eyes. Eye Vis (Lond) 2022;9:21.
crossref pmid pmc pdf
31. Kiraly L, Stange J, Kunert KS, Sel S. Repeatability and agreement of central corneal thickness and keratometry measurements between four different devices. J Ophthalmol 2017;2017:6181405.
crossref pmid pmc pdf
32. Song H, Liu C, Yang W, et al. Comparison of central corneal thickness measured in myopic eyes by Pentacam, Sirius and IOLMaster 700. Photodiagnosis Photodyn Ther 2024;49:104302.
crossref pmid
33. Saad A, Steinberg J, Frings A. Repeatability of pupil diameter measurements using three different topography devices. PLoS One 2023;18:e0290417..
crossref pmid pmc
34. Fang X, Xue W, Yu C, et al. Correlation between pupillary size and depth of focus after the implantation of extended depth of focus intraocular lenses. Graefes Arch Clin Exp Ophthalmol 2024;262:3897-903.
crossref pmid pdf
35. Tana-Rivero P, Aguilar-Corcoles S, Rodriguez-Prats JL, et al. Agreement of white-to-white measurements with swept-source OCT, Scheimpf lug and color LED devices. Int Ophthalmol 2021;41:57-65.
crossref pmid pdf
36. Vanathi M. Lens sizing calculation in phakic lens implantation: what is the best applicable measurement? Indian J Ophthalmol 2024;72:923-4.
crossref pmid pmc

Fig. 1
Bland-Altman plots comparing measurements of mean anterior keratometry (Km) and total keratometry (TKm) obtained with Anterion (Heidelberg Engineering Inc.), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec). For Bland-Altman analysis, the difference was defined as device 2 - device 1. (A) Km in Anterion vs. Pentacam HR. (B) Km in Anterion vs. IOLMaster 700. (C) Km in Pentacam HR vs. IOLMaster 700. (D) TKm in Anterion vs. Pentacam HR. (E) TKm in Anterion vs. IOLMaster 700. (F) TKm in Pentacam HR vs. IOLMaster 700. Solid black line indicates the mean difference; dotted lines indicate the 95% limits of agreement. SD = standard deviation; D = diopters.
kjo-2025-0202f1.jpg
Fig. 2
Deming regression plots showing measurements of mean anterior keratometry (Km) and total keratometry (TKm) obtained with Anterion (Heidelberg Engineering Inc.), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec). (A) Km in Anterion vs. Pentacam HR. (B) Km in Anterion vs. IOLMaster 700. (C) Km in Pentacam HR vs. IOLMaster 700. (D) Mean total corneal power (TCPm) in Anterion vs. mean total corneal refractive power (TCRPm) in Pentacam HR. (E) TCPm in Anterion vs. TKm in IOLMaster 700. (F) TCRPm in Pentacam HR vs. TKm in IOLMaster 700. The solid black line indicates the Deming regression line; the dotted line represents the line of identity (y = x). Estimated slope and intercept values with 95% confidence intervals are provided in each panel. D = diopters.
kjo-2025-0202f2.jpg
Fig. 3
Bland-Altman plots comparing central corneal thickness (CCT), pupil diameter (PD), and white-to-white (WTW) values obtained with Anterion (Heidelberg Engineering Inc.), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec). For Bland-Altman analysis, the difference was defined as device 2 - device 1. (A) CCT in Anterion vs. Pentacam HR. (B) CCT in Anterion vs. IOLMaster 700. (C) CCT in Pentacam HR vs. IOLMaster 700. (D) PD in Anterion vs. Pentacam HR. (E) PD in Anterion vs. IOLMaster 700. (F) PD in Pentacam HR vs. IOLMaster 700. (G) WTW in Anterion vs. IOLMaster 700. Solid black line indicates the mean difference; dotted lines indicate the 95% limits of agreement. SD = standard deviation.
kjo-2025-0202f3.jpg
Fig. 4
Deming regression plots for central corneal thickness (CCT), pupil diameter (PD), and white-to-white (WTW) obtained with Anterion (Heidelberg Engineering Inc.), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec). (A) CCT in Anterion vs. Pentacam HR. (B) CCT in Anterion vs. IOLMaster 700. (C) CCT in Pentacam HR vs. IOLMaster 700. (D) PD in Anterion vs. Pentacam HR. (E) PD in Anterion vs. IOLMaster 700. (F) PD in Pentacam HR vs. IOLMaster 700. (G) WTW in Anterion vs. IOLMaster 700. The solid black line indicates the Deming regression line; the dotted line represents the line of identity (y = x). Estimated slope and intercept values with 95% confidence intervals are provided in each panel.
kjo-2025-0202f4.jpg
Table 1
Mean anterior segment parameters measured by three devices
Parameter Anterion Pentacam HR IOLMaster 700 p-value
Kf (D) 43.59 ± 1.46 43.64 ± 1.48 43.68 ± 1.50 0.010*
Ks (D) 44.38 ± 1.48 44.42 ± 1.49 44.58 ± 1.55 <0.001*
Km (D) 43.98 ± 1.44 44.02 ± 1.46 44.12 ± 1.49 0.009*
TKf (D) 43.07 ± 1.53 43.39 ± 1.60 43.72 ± 1.56 <0.001*
TKs (D) 43.90 ± 1.52 44.22 ± 1.62 44.69 ± 1.57 <0.001*
TKm (D) 43.48 ± 1.50 43.81 ± 1.58 44.20 ± 1.53 <0.001*
CCT (μm) 541.66 ± 40.52 549.67 ± 39.62 542.73 ± 42.33 <0.001*
PD (mm) 6.62 ± 1.00 5.66 ± 1.00 6.63 ± 0.98 <0.001*
WTW (mm) 11.60 ± 0.40 - 11.80 ± 0.37 <0.001*

Values are presented as mean ± standard deviation. Measurements were obtained using Anterion (Heidelberg Engineering Inc.), Pentacam HR (Oculus), and IOLMaster 700 (Carl Zeiss Meditec). ANOVA p-values indicate overall differences among devices; for WTW, a paired t-test was used (Anterion vs. IOLMaster 700). TK from the Anterion = total corneal power (TCP), TK from the Pentacam HR = total corneal refractive power (TCRP), and TK from the IOLMaster 700 was adopted as provided by the device.

ANOVA = analysis of variance; Kf = flat keratometry; Ks = steep keratometry; Km = mean keratometry; TKf = flat total keratometry; TKs = steep total keratometry; TKm = mean total keratometry; CCT = central corneal thickness; PD = pupil diameter; WTW = white-to-white.

* Statistically significant (p < 0.05).

Table 2
Pairwise post-hoc comparisons of ocular parameters between devices
Variable/device comparison Mean difference 95% CI p-value
Mean anterior keratometry (D)
 Pentacam HR - Anterion 0.04 ± 0.03 −0.024 to 0.109 0.371
 IOLMaster 700 - Anterion 0.14 ± 0.02 0.096 to 0.190 <0.001*
 IOLMaster 700 - Pentacam HR 0.10 ± 0.03 0.040 to 0.161 <0.001*
Mean total keratometry (D)
 Pentacam HR - Anterion 0.33 ± 0.04 0.238 to 0.413 <0.001*
 IOLMaster 700 - Anterion 0.71 ± 0.02 0.662 to 0.760 <0.001*
 IOLMaster 700 - Pentacam HR 0.39 ± 0.04 0.300 to 0.472 <0.001*
Central corneal thickness (μm)
 Pentacam HR - Anterion 8.01 ± 0.58 6.596 to 9.416 <0.001*
 IOLMaster 700 - Anterion 1.06 ± 0.46 −0.049 to 2.174 0.066
 IOLMaster 700 - Pentacam HR −6.94 ± 0.71 −8.663 to −5.224 <0.001*
Pupil diameter (mm)
 Pentacam HR - Anterion −0.96 ± 0.07 −1.134 to −0.779 <0.001*
 IOLMaster 700 - Anterion 0.01 ± 0.05 −0.118 to 0.128 >0.999
 IOLMaster 700 - Pentacam HR 0.96 ± 0.06 0.810 to 1.114 <0.001*
White-to-white (mm)
 IOLMaster 700 - Anterion 0.19 ± 0.26 0.153 to 0.235 <0.001*,†

Values are shown as mean difference ± standard error for post hoc pairwise comparisons after repeated-measures ANOVA and as mean difference ± standard deviation for the paired WTW comparison, with corresponding 95% CI. Mean differences are defined as device 2 - device 1.

ANOVA = analysis of variance; CI = confidence interval; D = diopters.

* Statistically significant (p < 0.05);

† Paired t-test.

Table 3
Correlation (Pearson r) and agreement (ICC) between devices, including overall ICC across three devices (A-P-I)
Parameter Pearson r ICC


A - P A - I P - I A - P A - I P - I Overall (A - P - I)
Kf 0.963 0.969 0.959 0.963 0.967 0.958 0.963
Ks 0.969 0.984 0.970 0.969 0.974 0.964 0.969
Km 0.971 0.986 0.977 0.971 0.981 0.975 0.976
TKf 0.957 0.977 0.952 0.937 0.899 0.932 0.923
TKs 0.942 0.983 0.938 0.921 0.871 0.900 0.897
TKm 0.959 0.987 0.961 0.938 0.889 0.932 0.919
CCT 0.984 0.991 0.978 0.964 0.990 0.960 0.972
PD 0.590 0.800 0.693 0.530 0.406 0.801 0.471
WTW - 0.766 - - 0.677 - -

Pearson correlation coefficients (r) and ICC (two-way random, absolute agreement) with corresponding device pairs are presented. Overall ICC (A - P - I) represents the reliability across all three devices simultaneously. All correlation and ICC values were statistically significant at p < 0.001. Values close to 1.0 indicate stronger correlation or agreement: <0.50, poor reliability; 0.50 to <0.75, moderate reliability; 0.75 to 0.90, good reliability; and >0.90, excellent reliability.

ICC = intraclass correlation coefficient; A = Anterion (swept-source OCT, Heidelberg Engineering); P = Pentacam HR (Scheimpflug tomography, Oculus); I = IOLMaster 700 (swept-source optical biometer, Carl Zeiss Meditec); Kf = flat keratometry; Ks = steep keratometry; Km = mean keratometry; TKf = flat total keratometry; TKs = steep total keratometry; TKm = mean total keratometry; CCT = central corneal thickness; PD = pupil diameter; WTW = white-to-white.

Table 4
Inter-device differences in simulated IOL power based on total keratometry
Comparison Anterion - IOLM Pentacam HR - IOLM Pentacam HR - Anterion
Mean difference (D) 0.87 ± 0.39 0.48 ± 0.59 −0.39 ± 0.62
p-value <0.001 <0.001 <0.001
Diff. <0.5 D (%) 2.8 23.1 29.4
Diff. 0.5 to <1.0 D (%) 33.6 44.1 44.8
Diff. ≥1.0 D (%) 63.6 32.9 25.9

Values are shown as mean difference ± standard deviation. Absolute differences (Diff.) were categorized into three categories (<0.5 D, no change in lens power; 0.5 to <1.0 D, one-step difference; ≥1.0 D, two or more step differences). Simulated clinical impact of inter-device differences in total keratometry on simulated IOL power calculation using the Barrett Universal II formula. Pairwise differences in predicted IOL power were calculated using identical biometric inputs except for total keratometry values from each device. Mean differences are defined as (first device - second device). p-values were obtained using paired t-tests.

IOL = intraocular lens; IOLM = IOLMaster 700; D = diopters.