A popular device for noncontact ocular biometry is the IOLMaster 500 (Carl Zeiss AG, Jena, Germany), which was introduced in 1999 using partial coherence interferometry. This device can measure ocular biometric measurements such as AL, steep and flat K, ACD, white-to-white, and pupil diameter, and shows good agreement with the contact method (A-scan) using ultrasound [
2,
3]. After that, other optical devices using different principles have also been introduced to the market and have shown comparable or superior accuracy with the IOLMaster 500. The Lenstar LS 900 (Haag-Streit AG, Koeniz, Switzerland), which uses optical low-coherence reflectometry, and the OA-2000 (Tomey, Nagoya, Japan) [
4], which uses low coherence interferometer and swept-source combined with a placido disc topography are widely used these days. Recently, the efficacy of ANTERION (Heidelberg Engineering, Heidelberg, Germany), which is equipped with two imaging devices, swept-source optical coherence tomography, and an infrared camera, has been reported and its compatibility with existing devices has been demonstrated [
5-
9]. However, no study has compared ANTERION with the OA-2000 biometer so far. Therefore, this study aimed to measure and compare ocular biometric values using ANTERION with OA-2000 and IOLMaster 500 in cataract patients. Furthermore, we compared the predicted intraocular lens power with various intraocular lens power calculating formulas using the three devices.
Results
A total of 54 participants were enrolled in this study; however, three were excluded from the study as the AL could not be measured using the IOLMaster 500 due to posterior subcapsular cataract (NO2NC2C1P2, NO3NC3C2P3, and NO2NC2C2P3 based on Lens Opacities Classification System III grade [
25]) in the center of the lens. Finally, 51 eyes from 51 participants were analyzed. The mean age of the participants was 67.2 years (range, 45 to 79 years). Among the total participants, 18 participants were male and 33 participants were female.
The average flat K measured using the ANTERION, the OA-2000, the IOLMaster 500, and the AR/K device was 43.58 ± 1.39, 43.83 ± 1.39, 43.93 ± 1.47, and 43.87 ± 1.46 D, respectively (
Table 1). ANTERION yielded the flattest values, and there were significant differences in the values measured using ANTERION and the other devices (
p < 0.05) (
Table 2). Analysis of agreement indicated that ANTERION and the OA-2000 were interchangeable (95% LoA, 0.86 D) (
Table 2 and
Fig. 1A-1C). On the other hand, the IOLMaster 500 was shown to be not interchangeable with both ANTERION (95% LoA, 1.18 D) and the OA-2000 (95% LoA, 1.06 D). When comparing flat K values measured by the three devices with that measured by conventional AR/K device, all three optical biometers showed good correlations with the AR/K device (all
p > 0.9), however, ANTERION showed significantly flatter flat K than that of the AR/K device (
p < 0.001). In agreement analysis for flat K, all optical biometers were not considered interchangeable with the AR/K device (95% LoA ≥1.25 D) (
Table 2).
The mean steep K measured using the ANTERION, the OA-2000, the IOLMaster 500, and the AR/K device was 44.45 ± 1.22, 44.77 ± 1.29, 44.85 ± 1.30, and 44.66 ± 1.30 D, respectively (
Table 1). The value measured by ANTERION was the flattest, and there were significant differences in the values measured using ANTERION and the other devices (
p < 0.05) (
Table 2). Meanwhile, there was no significant difference between the steep K measured using the IOLMaster 500 and OA-2000 (
p = 0.053). In terms of agreement, the OA-2000 was shown to be interchangeable with both ANTERION (95% LoA, 0.93 D) and the IOLMaster 500 (95% LoA, 0.93 D) (
Table 2 and
Fig. 1D-1F). When comparing the steep K values with that measured by the AR/K device, ANTREION showed significantly flatter steep K (
p < 0.001) and IOLMaster 500 showed significantly steeper steep K than the AR/K device (
p = 0.001) (
Table 1). In an agreement analysis, steep K values measured by three optical biometers were not interchangeable with that measured by the AR/K device (95% LoA ≥1.09 D) (
Table 2).
Regarding corneal astigmatism, the mean
J0 vector component of astigmatism obtained from the ANTERION, the OA-2000, the IOLMaster 500, and the AR/K device was 0.06 ± 0.39, −0.09 ± 0.37, 0.04 ± 0.42, and 0.02 ± 0.34 D, respectively. The mean
J45 vector component of astigmatism obtained from the ANTERION, the OA-2000, the IOLMaster 500, and the AR/K device was −0.07 ± 0.36, −0.01 ± 0.43, 0.06 ± 0.35, and −0.04 ± 0.34, respectively (
Table 1). There were no significant differences in the
J0 (
p = 0.200) and
J45 (
p = 0.307) vector components of astigmatism between the three devices (
Table 1). However, the agreement analysis between the three devices revealed to be poor (95% LoA ≥1.97 D) (
Table 2 and
Fig. 2A-2F). All optical biometers showed no significant difference in
J0 and
J45 with the AR/K device (all
p > 0.05) yet showed poor agreements with the AR/K device (95% LoA ≥1.83 D) (
Table 2).
The mean ACD measured using the ANTERION, the OA-2000, and the IOLMaster 500 was 3.08 ± 0.38, 3.05 ± 0.74, and 3.13 ± 0.39, respectively (
Table 1). There were significant differences among the measurements yielded by all devices (all
p < 0.05) (
Table 3). Good agreement was observed only between ANTERION and the OA-2000 (95% LoA, 0.15 mm) (
Table 3 and
Fig. 3A-3C).
The mean AL measured using ANTERION, the OA-2000, and the IOLMaster 500 was 23.64 ± 1.33, 23.64 ± 1.32, and 23.61 ± 1.32 mm, respectively (
Table 1). There was no significant difference in the mean AL between the ANTERION and OA-2000 (
p = 0.917). The AL of the IOLMaster 500 significantly differed from that of the other devices (all
p < 0.05) (
Table 3). However, the difference was not clinically significant (range, 0.01 to 0.03 mm). Good agreement was observed among all devices with a narrow 95% LoA (≤0.17 mm) (
Table 3 and
Fig. 4A-4C).
Predicted intraocular lens powers, which becomes emmetropic using the SRK/T, Haigis, Barrett Universal II, and Kane formulas were compared (
Table 4). The predicted intraocular lens power calculated by four formulas did not significantly differ between the OA-2000 and IOLMaster 500 (
p > 0.999); however, the 95% LoA range was wide (1.24 to 1.70 D). The predicted intraocular lens power measured using ANTERION was approximately 0.34 to 0.40 D higher than that measured using the other devices (
p < 0.001) (
Table 5).
Discussion
This study compared ocular biometrics measured using the newest swept-source optical coherence tomography-based optical biometer, the ANTERION device, and the existing OA-2000 and IOLMaster 500 devices. The flat K and steep K were the flattest when measured by ANTERION, and the ACD was the deepest when measured by the IOLMaster 500. There was a high correlation between the measurements of all devices except the vector component of astigmatism; however, only the AL showed good agreement among all three devices. The predicted intraocular lens power, which was calculated using the SRK/T, Haigis, Barrett Universal II, and Kane formulas, did not significantly differ for all formulas between the IOLMaster 500 and OA-2000; however, the agreement was low. In addition, ANTERION demonstrated a higher predicted intraocular lens power than the other two devices.
In previous studies comparing OA-2000 and IOLMaster 500, mean K (95% LoA, −0.59 to +0.36 [
26] or −0.97 to +1.03 D [
27]), ACD (95% LoA, −0.58 to +0.39 [
26] or −0.33 to +0.54 mm [
27]), and AL (95% LoA, −0.16 to +0.05 [
26] or −0.09 to +0.1 mm [
27]) showed good agreement in 102 eyes of 68 patients with cataracts [
26] and in 58 eyes of 58 patients with cataracts [
27]. In contrast, studies comparing ANTERION and the IOLMaster 500 reported inconsistent agreements according to the studies. In a study of 48 eyes of 48 adults with or without cataracts, Chan et al. [
22] compared the mean keratometry, vector components of astigmatism (
J0 and
J45), central corneal thickness, ACD, and AL. They reported that there were poor agreements in the mean keratometry,
J0,
J45, and the ACD: the 95% LoA for the mean keratometry,
J0,
J45, ACD, and AL was −0.588 to +0.357, −0.346 to +0.651, −0.398 to +0.485 D, −0.260 to +0.263, and −0.055 to +0.013 mm, respectively. In contrast, when comparing ANTERION, the MS-39 (Costruzione Strumenti Oftalmici, Florence, Italy), and the IOLMaster 500 in 96 eyes of 96 adults with or without cataracts, ANTERION and the MS-39 showed high agreement for simulated keratometry, posterior keratometry, central corneal thickness, and aqueous depth; and ANTERION and the IOLMaster 500 showed high agreement for simulated keratometry (95% LoA, −0.68 to +0.70 D), ACD (95% LoA, −0.50 to +0.57 mm), AL (95% LoA, −0.06 to +0.05 mm), and corneal diameter (95% LoA, −0.72 to +0.31 mm) [
7]. Kim et al. [
12] also reported that the flat K, steep K, J0, J45, ACD and AL exhibited good agreement (95% LoA, −1.18 to +0.83, −1.06 to +0.95, −0.60 to +0.73, −0.40 to +0.50 D, −0.14 to +0.26, and −0.15 to +0.21 mm, respectively) between ANTERION and the IOLMaster 500 in 175 eyes of 107 subjects aged 40 years or younger with clear lens or mild cataracts. One of the reasons for the discrepancies in previous studies may be because these previous studies interpreted the findings without defining the range of agreement. If clinical agreement for intraocular lens power calculation is defined as good agreement for 95% LoA of corneal curvature within 1.0 D when comparing ANTERION and IOLMaster 500, the range of 95% LoA in the previous study that reported high agreement (1.38 D [
7]) can lead to an interpretation of low agreement in line with this study (95% LoA range steep K to flat K, 1.20 to 1.18 D). Likewise, the range of 95% LoA of flat K and steep K (2.01 and 2.01 D, respectively) in the study of Kim et al. [
12] can be interpreted as low agreement.
A comparison of the ANTERION and the OA-2000 showed flatter corneal curvature for the ANTERION; however, the Bland-Altman plot showed good agreement between the two devices (95% LoA ≤0.93 D). The 95% LoA between the IOLMaster 500 and OA-2000 were 0.93 D for the steep K and 1.06 D for the flat K. The K between the two devices showed good or low agreement depending on the curvature. These differences may be caused by the different principles and ranges of K for each device.
In relation to
J0 and
J45 vector components of astigmatism, this study revealed a wider 95% LoA range (2.51 and 1.97 D, respectively) than previous comparison studies (
J0 95% LoA range, 0.997 [
22] and 1.33 D [
12];
J45 95% LoA range, 0.883 [
22] and 0.90 D [
12]) comparing the ANTERION with the IOLMaster 500. The agreements for the
J0 and
J45 vector components measured using the three devices (1.83 ≤ 95% LoA ≤ 2.52) were relatively poorer than that for cylinder power (1.10 ≤ 95% LoA ≤ 1.53) (
Table 2 and
Supplementary Table 1). Thus, the lower agreement of devices regarding astigmatic vector components could be due to the corneal cylinder orientation rather than cylinder power. These findings might have occurred because of the relatively poor repeatability and reliability of the axis measurements performed using the devices [
15,
28]. This outcome should be considered when planning cataract surgery with toric intraocular lens implantation, considering preoperative astigmatism and postoperative residual astigmatism would affect visual function and vision-related quality of life [
29].
For ACD, a measurement error of 0.1 mm causes a refractive error of 0.15 D [
30], and a 95% LoA range corresponding to 20% of the mean ACD may be too broad for clinical purposes such as intraocular lens power calculation [
20]. In this study, the ACD measured using the IOLMaster 500 was significantly deeper than that measured using ANTERION or the OA-2000 and showed low agreement (95% LoA ≥0.55 mm). Additionally, the ANTERION and the OA-2000 (95% LoA, 0.15 mm) showed good agreement for the measurement of ACD. Using the same criteria (95% LoA <0.4 mm) as our study, the results of all previous studies comparing ACD between the IOLMaster 500 and ANTERION (95% LoA, 0.4 [
12], 0.52 [
22], and 1.07 mm [
7]) or the IOLMaster 500 and the OA-2000 (95% LoA, 0.87 [
27] and 0.97 mm [
26]) can be interpreted as low agreement, as observed in this study. As suggested by Elbaz et al. [
20] and Chan et al. [
22], the IOLMaster 500 cannot measure the axial ACD, as this device measures the ACD with a slit source projected from the temporal side, which would have led to deeper ACD measurements compared to other devices.
In this study, AL was the only variable that showed a high correlation and good agreement among all devices (all 95% LoA ≤0.17 mm). The IOLMaster 500 and OA-2000 showed the narrowest 95% LoA (0.08 mm) and the highest agreement. This was a narrower range than that reported in previous studies (0.21 [
26] and 0.19 mm [
27]). Among the studies that compared ANTERION and the IOLMaster 500, the range of 95% LoA was the widest (0.36 mm) in the study by Kim et al. [
12], and this may be attributed to the participants of that study, which included children aged five years or older with poor cooperation.
The major limitation of the IOLMaster 500 is its high failure rate (38%) in the acquisition of AL in dense cataract or posterior subcapsular cataracts [
31]. In a study by McAlinden et al. [
23], the IOLMaster 500 showed a failure rate of 36.07%, whereas the OA-2000 demonstrated a failure rate of 0%. Three participants in this study were excluded from the analysis because the AL could not be measured using IOLMaster 500 due to posterior subcapsular cataract, whereas the AL of these three patients could be measured using the OA-2000 and ANTERION. This may be attributed to the longer wavelength of the OA-2000 and ANTERION (1,060 and 1,300 nm, respectively) for the measurement of AL compared with the IOLMaster 500 (780 nm), which increases penetration [
23].
As for intraocular lens power calculation, no other study has compared the predicted intraocular lens power between ANTERION and other devices to date. Shetty et al. [
21,
32] reported a high intraclass correlation coefficient (≥0.977) of predicted intraocular lens powers in ANTERION, Lenstar LS 900, and IOLMaster 700 (Carl Zeiss AG). Kongsap [
26] assessed intraocular lens power using the SRK/T formula derived from the OA-2000 and the IOLMaster 500 and reported similar intraocular lens power derived from both instruments (mean difference of 0.32 D), with an excellent correlation (r = 0.989) in cataract patients. In this study, various intraocular lens formulas (SRK/T, which is a third-generation formula [
33], Haigis, which is a fourth generation formula [
2], and the Barrett Universal II [
34] and Kane formulas [
35], which are fifth generation formulas with very high accuracy that have been introduced most recently) were used to calculate the predicted intraocular lens power in the three devices. Regardless of the formula, the predicted intraocular lens powers of ANTERION were significantly higher (0.34 and 0.40 D, respectively) than that of the other devices (
p < 0.001). Additionally, agreement analysis revealed that ANTERION is clinically not interchangeable with other devices (95% LoA ≥1.04 D). Even though the power of most intraocular lens is 0.5 D increments, this difference may cause the postoperative refractive error to be more myopic. There was no significant difference in the predicted intraocular lens power between the IOLMaster 500 and OA-2000; however, the 95% LoA range was also wide (≥1.04 D). Considering that every 1.0 D error of the predicted intraocular lens power leads to 0.67 D of refractive errors [
1], it would be difficult to conclude that the calculated intraocular lens powers of the devices are interchangeable.
The limitations of this study are as follows. The sample size was small, and those with a flat K (<41.0 D), steep K (>48.5 D), short AL (<22.0 mm), and long AL (>26.0 mm) were not included. It will be necessary to study such patients as well as a larger number of participants with various conditions, such as patients with corneal opacity or postrefractive surgery patients. In addition, analyzing the difference between the predicted target refraction using the predicted intraocular lens power and the postoperative measured refractive error would provide additional information for the comparison of ocular biometric measurements.
In conclusion, this study is the first to compare ocular biometrics measured using ANTERION, OA-2000, and IOLMaster 500 in eyes with cataracts. Each biometric differed significantly according to the device used. The flat K, steep K, J0, J45, and ACD, except AL, exhibited low agreement among the devices. The low agreements were also observed in the predicted intraocular lens power among the devices.